Aution Max is the highly evolved member of the Aution Orchestra family of urine analysers by A. Menarini Diagnostics. The ideal assistant for optimising cost and time efficiency in medium-to-large sized laboratories. The total diagnostics system featured by Aution Max develops around the heart of this technological genius. With the capacity of automatically managing advanced and heavy-duty urinalysis routines, Aution Max can work in full online/offline connection with a dedicated back-up environment. Measurements are performed via the accurate control & check system, supported by both digital codes sample recognition and the 1000 result capacity of the memory storage. Aution Max, competence and mastery by A. Menarini Diagnostics.
WHEN AUTION MAX IS IN ACTION,THE JOB IS JUST DONE.
|
|
|
- Fully automated operation
- Bar code sample control
- 1000 results memory capacity
- Sample dropping method (minimum amount of urine only 2 mg)
- Abnormal urine colour detection
- Colour-tone & turbidity measurement
- Automatic temperature compensation
- maximum sample measurement relibility with carry-over effect control
|
|
"Evaluation of Aution Max AX-4280 Automated Urine Test-Strip Analyser" by Zahur Zaman et al.Clin Chem Lab Med 2001; 39(7): from p.649 to p.657
Specifications:
|
Sample
|
Urine
|
|
Meas. Item
|
GLU, KET, BIL, NIT, PRO, Ph, BLD, LEU, Specific Gravity, Colour tone, Turbidity
|
|
Meas. Wavelengths
|
430, 500, 565, 635, 760 nm
|
|
Meas. Method
|
Test strip; Dual (Single) wavelength reflectance method
Specific gravity:Refractive index method
Colouration: Four-wavelength reflectance method Turbidity: Transparency index method
|
|
Meas. Range
|
Test strip: Table on page
Specific gravity: 1.000-1.050 (Resolution: 0.0001)
Colouration: 23 different colour-tone simultaneously Turbidity: Clear, Turbid or Excessively turbid
|
|
Test Strip
|
Uriflet S 9UB
|
|
Test Strip Reaction Time
|
Approx. 60 sec.
|
|
Measuring Time
|
225 samples/hour
|
|
Max. Loading Strip
|
200 strips
|
|
Required Sample Vol.
|
2.0 ml
|
|
Absorbing Sample Vol.
|
0.95 ml
|
|
Sample Container
|
Urine sample tube
|
|
Max. Loading Sample
|
100 samples
|
|
Memory Capacity
|
1000 tests (including 180 control Measurement results)
|
|
Display
|
20 digits X2 lines LCD
|
|
Built-in Printer
|
36-character thermal printer (58 mm width)
|
|
External Output
|
RS-232C interface (Control by XON/XOFF is available)
|
|
Transmission Rate
|
300, 600, 1200, 2400, 4800, 9600 bps
|
|
Operating Condition
|
Temperature: 10°-30° C /Humidity: 20-80%
|
|
Meas. Condition
|
Temperature: 10°-30° C (with temperature compensation function) Humidity: 30-60%
|
 |
|
Power Supply
|
AC 100/120/220/230 V, 50/60 Hz
|
|
Power Input
|
Max. 250 W
|
|
Dimension
|
Analyser: 530 (W) X 330 (D) X 530 (H) mm Sampler: 530 (W) X 200 (D) X 130 (H) mm
|
|
Weight
|
Analyser: 33 Kg, Sampler: 4 Kg
|
Aution Max strips:
|
|
|
|
|
|
|
|