Menarini / Diagnostic / Products / Urinalysis / Aution Max
 
Italiano - Contact us - Sitemap
 Italy
Diagnostics Italy
Aution Max
 

Aution Max is the highly evolved member of the Aution Orchestra family of urine analysers by A. Menarini Diagnostics. The ideal assistant for optimising cost and time efficiency in medium-to-large sized laboratories. The total diagnostics system featured by Aution Max develops around the heart of this technological genius. With the capacity of automatically managing advanced and heavy-duty urinalysis routines, Aution Max can work in full online/offline connection with a dedicated back-up environment. Measurements are performed via the accurate control & check system, supported by both digital codes sample recognition and the 1000 result capacity of the memory storage. Aution Max, competence and mastery by A. Menarini Diagnostics.

WHEN AUTION MAX IS IN ACTION,THE JOB IS JUST DONE.

  • Fully automated operation
  • Bar code sample control
  • 1000 results memory capacity
  • Sample dropping method (minimum amount of urine only 2 mg)
  • Abnormal urine colour detection
  • Colour-tone & turbidity measurement
  • Automatic temperature compensation
  • maximum sample measurement relibility with carry-over effect control

"Evaluation of Aution Max AX-4280 Automated Urine Test-Strip Analyser" by Zahur Zaman et al.Clin Chem Lab Med 2001; 39(7): from p.649 to p.657

Specifications:

Sample

Urine

Meas. Item

GLU, KET, BIL, NIT, PRO, Ph, BLD, LEU, Specific Gravity, Colour tone, Turbidity

Meas. Wavelengths

430, 500, 565, 635, 760 nm

Meas. Method

Test strip; Dual (Single) wavelength reflectance method
Specific gravity:Refractive index method
Colouration: Four-wavelength reflectance method
Turbidity: Transparency index method

Meas. Range

Test strip: Table on page
Specific gravity: 1.000-1.050 (Resolution: 0.0001)
Colouration: 23 different colour-tone simultaneously
Turbidity: Clear, Turbid or Excessively turbid

Test Strip

Uriflet S 9UB

Test Strip Reaction Time

Approx. 60 sec.

Measuring Time

225 samples/hour

Max. Loading Strip

200 strips

Required Sample Vol.

2.0 ml

Absorbing Sample Vol.

0.95 ml

Sample Container

Urine sample tube

Max. Loading Sample

100 samples

Memory Capacity

1000 tests (including 180 control Measurement results)

Display

20 digits X2 lines LCD

Built-in Printer

36-character thermal printer (58 mm width)

External Output

RS-232C interface (Control by XON/XOFF is available)

Transmission Rate

300, 600, 1200, 2400, 4800, 9600 bps

Operating Condition

Temperature: 10°-30° C /Humidity: 20-80%

Meas. Condition

Temperature: 10°-30° C (with temperature compensation function)
Humidity: 30-60%

Power Supply

AC 100/120/220/230 V, 50/60 Hz

Power Input

Max. 250 W

Dimension

Analyser: 530 (W) X 330 (D) X 530 (H) mm
Sampler: 530 (W) X 200 (D) X 130 (H) mm

Weight

Analyser: 33 Kg, Sampler: 4 Kg

Aution Max strips:

Aution Max Specifications