Menarini / Diagnostic / Products / Haemoglobin Analyser / System specifications
 
Contact us - Sitemap
Greek - English
Greece
System specifications

Samples:

Whole blood and/or Haemolysate

Measured Items:

Diabetes Monitoring Mode: HbA1c, HbA1, HbF, Main Variant Hbs Detection
Diabetes Monitoring & Thalassemia Screening Mode: HbA1c, HbA1, HbF,HbA2, Main Variant Hbs Quantification

Measurement Principle:

Reversed-phase cation exchange chromatography

Detection Method:

Dual-wavelength colorimetry (415-500 nm)

Resolution:

0.1%

Throughput:

Diabetes Monitoring Mode: 2.9 minutes per sample Diabetes Monitoring & Thalassemia Screening Mode: 4.2 minutes per sample

Sample Volume:

Approximately 4 µL whole blood

Sample Tubes:

Blood Collection Tube 12,3 to 15 diameter x 75 to 100 mm length Sample Cup 500 µL

Sampling method:

Cap Piercing

Rack:

Sysmex compatible

Column Temperature:

Approximately 40° C

External Output:

RS-232C (2-way), 1 channel, 300 to 9600 bps

Operating Temperature:

10 to 30° C

Operating Humidity:

20 to 80% (no condensation)

Display:

2-row x 20-digit LCD

Built-in Printer:

58 mm wide thermal printer

Sample Loading Capacity:

A maximum of 100 samples. Continuous loading possibility

Min. Required Sample Volume:

Blood Collection Tube: 1 mL Sample Cup: 400 µL

Power Supply:

200 V AC, 50/60 Hz

Power Input:

320 VA

External Dimensions:

530W x 530D x 530H mm

Weight:

Approximately 43 Kg